2 Technical data
Temperature ranges: IN 5: 0 ... 100°C 0 ... 500°C IN 5/5: 100 ... 600°C
0 ... 200°C 0 ... 900°C 200 ... 800°C
0 ... 300°C -32 ... 50°C 100 ... 1300°C
0 ... 400°C -32 ... 900°C 400 ... 2500°C
(other ranges for both instruments on request)
Data handling: Digital
Spectral range: IN 5: 8...14 µm IN 5/5: 5.14 µm
Lens: IN 5: Germanium (Ge) IN 5/5: Zinc Sulfide (ZnS)
IR-Detektor: Thermopile
Power supply: 24 V DC (10... 30 V) nominal, ripple must be less than 0.5 V
Analog output: 4 ... 20 mA direct current, linear resolution: 2880 steps
Load:
max. 700 Ω at 24 V (max. 100 Ω at 12 V)
Emissivity: 0.2 ... 1 switchable in the instrument in steps of 0.01
Exposure time t
90
: 0.08 s; adjustable to 0.5 s; 1 s; 2 s; 5 s
Measurement uncertainty:
dependent on object tempera-
ture T and ambient tempera-
ture T
amb
(ε = 1, t
90
= 1 sec):
IN 5:
T= -32...0°C: 1.5°C (T
amb
= 15...30°C); 2°C (T
amb
= 0…15 or 30…63°C)
T=0...300°C: 0.6% of reading in °C or 1°C (T
amb
= 15...30°C)
1)
1% of reading in °C or 1.5°C (T
amb
= 0...15 or 30...63°C)
1)
T=300...900°C: 1% of reading in °C (T
amb
= 15...30°C)
1.3 % of reading in °C (T
amb
= 0...15 or 30…63°C)
IN 5/5:
T < 1300°C: 0.6% of reading in °C or 2°C (T
amb
=15...30°C)
1)
1% of reading in °C or 1,5°C (T
amb
=0...15 or 30...63°C)
1)
T=1300...1800°C: 0.8% of reading in °C (T
amb
=15...30°C)
1.2% of reading in °C (T
amb
=0...15 or 30...63°C)
T=1800...2500°C: 1% of reading in °C (T
amb
=15...30°C)
1.4 % of reading in °C (T
amb
=0...15 or 30...63°C)
Repeatability (ε = 1, t
90
= 1 s):
0.3 % of measured value in °C or 0.6 °C
1)
Noise Equivalent Tempera-
ture Difference (NETD):
(ε = 1, T
amb
= 23°C)
IN 5: at t
90
= 80 ms: 0.2°C (at 23°C measuring temperature)
at t
90
= 1 s: 0.05°C (at 23°C measuring temperature)
IN 5/5: at t
90
= 80 ms: 0.7°C (at 110°C measuring temperature)
at t
90
= 1 s: 0.4°C (at 110°C measuring temperature)
Ambient temperature: 0 ... 70°C
0 ... 63°C for IN 5/5, MB 25
Storage temperature: -20 ... 70°C
Relative humidity Non condensing conditions
Protection class: IP65 (DIN 40050)
Weight: 410 g
Housing: Stainless steel
Operating position: Any
Dimensions:
CE-label: according to EU directives about electromagnetic immunity
1)
Whichever value is greater. The instrument must be at a constant ambient temperature for a minimum of 15 minutes