Lumasense-technologies M390 Bedienungsanleitung

Stöbern Sie online oder laden Sie Bedienungsanleitung nach Werkzeug Lumasense-technologies M390 herunter. LumaSense Technologies M390 User Manual Benutzerhandbuch

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken

Inhaltsverzeichnis

Seite 1 - Blackbody Calibration Source

MANUA LM390Blackbody Calibration Source

Seite 2

M390 Manual Introduction  102.2 Technical DataThisModelModel Part No. Exit Aperture Temperature Range Power[ ] M390-A1 14029-A1 0.625 in 600-

Seite 3 - Contents

M390 Manual Installation  113 Installation3.1 Installing the M390To install the M390:1. Allow the M390 to warm-up to room temperature for at least 2

Seite 4 - M390 Manual Contents  iv

M390 Manual Installation  12To ensure consistent document formatting, this page was intentionally left blank.

Seite 5 - 1 General Information

M390 Manual Operation  134 Operation4.1 Starting UpTo start up the M390:1. Turn on the power switch/breaker and allow 20 seconds for the system to s

Seite 6 - 1.4 Unpacking the Instrument

M390 Manual Operation  14Eurotherm controller4.2.1 Other Controller ModesIf, for some reason, the power must be turned off while monitoring temperat

Seite 7

M390 Manual Operation  158. Press the UP button to set new units °F (or K for Kelvin, if desired). Confirm the changewith the RETURN button or press

Seite 8

M390 Manual Operation  164.5.1 RS485 OptionRS485 two wire is available as an option. The connection scheme is shown below:RS485 for multi-drop appl

Seite 9 - 2 Introduction

M390 Manual Operation  174.6 PID Tuning ConstantsThe PID tuning constants have been factory set for optimum performance and are noted below.Heating

Seite 10 - 2.2 Technical Data

M390 Manual Operation  18Typical Radiance temperature calibration look-up table (LUT) used in our blackbodyTo gain access to the custom look-up tabl

Seite 11 - 3 Installation

M390 Manual Operation  19Note: Calibration points do not have to be used as pre-programmed. To dial in adifferent calibration point, simply change a

Seite 12

Confidential InformationThe material contained herein consists of information that is the property of LumaSenseTechnologies and intended solely for us

Seite 13 - 4 Operation

M390 Manual Operation  20To ensure consistent document formatting, this page was intentionally left blank.

Seite 14 - 4.3 Shutting Down

M390 Manual Maintenance  215 Maintenance5.1 Cavity LifeThe life expectancy of the graphite cavity depends on the time and temperature of operation.T

Seite 15 - M390 Manual Operation  15

M390 Manual Maintenance  225.4 Furnace Core DisassemblyThe graphite core can be replaced without disconnecting cooling or purge tubing. Refer toexpl

Seite 16 - 4.5.1 RS485 Option

M390 Manual Maintenance  235.5 Fuse LocationsThe M390 has several fuses. Refer to the table below. See the appendix for photo-view of fuselocations

Seite 17 - 4.7 Controller Calibration

M390 Manual Maintenance  245.6 Recommended Spare PartsModelTemperatureRange Cavity LengthReplacementCavity KitQuartz Tube, ElectrodeSeats, and O-Rin

Seite 18 - M390 Manual Operation  18

M390 Manual Troubleshooting  256 TroubleshootingSymptom Tests/ Solutions1. Blackbody Does not heat up Is there both Argon purge gas and coolingwater

Seite 19 - M390 Manual Operation  19

M390 Manual Troubleshooting  269. Blackbody heats up and is glowing, butthe controller does not indicate temperature> beginning of range.Check th

Seite 20 - M390 Manual Operation  20

M390 Manual Appendix A: Drawings  27Appendix A: Drawings

Seite 21 - 5 Maintenance

M390 Manual Appendix A: Drawings  28

Seite 22 - 5.4 Furnace Core Disassembly

M390 Manual Appendix B: ASCII/BASIC  29Appendix B: ASCII/BASICTypical ASCII request string to read the blackbody temperature:'-------make a req

Seite 23 - 5.5 Fuse Locations

M390 Manual Contents  iiiContents1 General Information ...

Seite 24 - 5.7 Cleaning

M390 Manual Appendix B: ASCII/BASIC  30A simple BASIC program to send blackbody setpoint:CLS'-----open comm port 1 for 9600 baud, 7,E,1-----OPE

Seite 25 - 6 Troubleshooting

M390 Manual Appendix C: Declaration of Conformance  31Appendix C: Declaration of Conformance

Seite 26

M390 Manual Appendix C: Declaration of Conformance  32To ensure consistent document formatting, this page was intentionally left blank.

Seite 27 - Appendix A: Drawings

M390 Manual Index  33IndexCCavity life 21Cleaning 24Controller Calibration 17Controller Modes 14DDisposal 7EEurotherm iTools Software 17FFurnace Cor

Seite 28

M390 Manual Index  34To ensure consistent document formatting, this page was intentionally left blank.

Seite 29 - Appendix B: ASCII/BASIC

M390 Manual Index  35

Seite 30

M390 Manual Contents  iv6 Troubleshooting ...

Seite 31

M390 Manual General Information  51 General Information1.1 Information about the user manualCongratulations on choosing the high quality and highly

Seite 32

M390 Manual General Information  6Warning: DO NOT USE NITROGEN (N2) FOR PURGE GAS. AT HIGH TEMPERATURES,CYNIDE GAS COULD BE GENERATED.1.3 Limit of L

Seite 33 - M390 Manual Index  33

M390 Manual General Information  7Return the instrument upon receipt of the RMA nµmber, transportation prepaid. Clearlyindicate the assigned RMA nµm

Seite 34 - M390 Manual Index  34

M390 Manual General Information  8To ensure consistent document formatting, this page was intentionally left blank.

Seite 35 - M390 Manual Index  35

M390 Manual Introduction  92 Introduction2.1 DescriptionThe M390 series are high temperature blackbody calibrators that can produce a very hightempe

Kommentare zu diesen Handbüchern

Keine Kommentare