IMPAC pyrometers ISR 12-LO · IGAR 12-LO
25
Reading emissivity: AAem Output: XXXX (dec. 0010 ... 1000 in ‰)
Emissivity: AAemXX XX = (10 ... 99%), 00 = 100% (decimal)
Emissivity: AAemXXXX XXXX = (0010 ... 1000‰) (decimal)
Emissivity slope: AAevXXXX XXXX = 0800 ... 1200 (equivalent to 0.8 ... 1.2)
Ratio part in
metal mode:
AAmvXX XX = 01 ... 99%
Exposure time t
90
: AAezX X = 0 ... 6 (decimal)
0 = intrinsic time constant of the device
1 = 0.01 s 3 = 0.25 s 5 = 3.00 s
2 = 0.05 s 4 = 1.00 s 6 = 10.00 s
Clear times of maxi-
mum value storage:
AAlzX
X = 0 ... 8 (dec.) 0 = Maximum value storage off
1 = 0.01 s 4 = 1.00 s 7 = external deletion
2 = 0.05 s 5 = 5.00 s 8 = automatically deletion
3 = 0.25 s 6 = 25.00 s 9 = hold
External deletion: AAlx
Simulation of an external deletion contact
Operation mode: AAkaX
X = 0 metal mode from channel 1 and 2
X = 1 mono mode
X = 2 ratio temperature
Switch-of limit: AAawXX XX = 02 ... 50%
Analog output: AAasX X = 0 or 1 0 = 0 to 20 mA 1 = 4 to 20 mA
Reading basic
temperature range:
AAmb Output: XXXXYYYY (hex 8-digit, °C or °F)
XXXX = beginning of temperature range
YYYY= end of temperature range
Reading temperature
sub range:
AAme same as mb
Setting of temperature
sub range:
AAm1XXXXYYYY XXXX (hex. 4-digit) beginning of temp. range (°C or °F)
YYYY (hex. 4-digit) end of temp. range (°C or °F)
Address: AAgaXX
XX = (00 ... 97)
00 ... 97 = regular device addresses
99 = global address with response
98 = global address without response (only setting commands!)
Baud rate: AAbrX X = 0...6 or 8 (dec.)
1 = 2400 baud 4 = 19200 baud (7 is not allowed)
2 = 4800 baud 5 = 38400 baud 8 = 115200 baud
3 = 9600 baud 6 = 57600 baud
Changing °C / °F AAfhX Output: X = 0: display in °C; X = 1: display in °F
Reading signal strength AAtr Output: XXXX, XXXX = 0000 ... 9999 in ‰
Warning level: AAdwXX XX = 00 ... 99% (2 digit, hex.)
Wait time: AAtwXX XX = 00 ... 99 (decimal)
Error status: AAfs Output: XX;
XX=00 … FF (00 = no error)
(01 … FF: error code for LumaSense service)
Internal temperature: AAgt
Output: XXX (dec. 000 ... 099°C or 032 ... 210°F)
Max. internal
temperature:
AAtm Output: XXX (dec. 000 ... 099°C or 032 ... 210°F)
Reading interface type: AAin
Output: X = 1 or 2; 1 = RS232, 2 = RS485
Laser targeting light: AAlaX
X = 0 switch off laser targeting light
X = 1 switch on laser targeting light
Lock keyboard: AAlkX
X = 0 ... 3
1 = lock lk1, removal with command lk0 or power off-on
0 = removal of lock lk1
3 = continuous lock lk3, removal only with command lk2
2 = removal of lock lk3
Kommentare zu diesen Handbüchern