Pearl Control Software Manual Using Calibrate Tab Features 38
8. Once a pixel has been replaced, the effect is only temporary unless it is saved to the
camera permanently. Click Save to Camera to save replaced pixels to the camera. This
operation will take several seconds to complete, and during this time the camera will
flicker.
Replace Column
Replaces the entire column with the selected replacement code.
To replace a column of pixels:
1. Click Help and select Advanced from the drop down menu.
2. Enter the password: engineer
3. Click on the Calibrate Tab.
4. Click on Pixel Replacement.
5. Select the pixel to replace by using the tools described in Section 4.4.2.